Polartronic H532 (Highest measuring precision)

  • Highest measuring precision and resolution (0.0001°) 
  • Measuring of optical rotatory dispersion (ORD) 
  • Up to 5 wavelengths
  • Calibration free because of spectral lamps

Spectral multi-wavelength polarimeter NEW – High resolution of 0.0001° !

The high-resolution (0.0001 °) polarimeter H532 with up to 5 wavelengths and the highest precision!

The technology of the Polartronic H532, which is securely packaged under the elegant housing, works with a further developed software, which is known from the ATR series refractometer, with a large programmable display. The input of the following is possible: sample identification numbers, statistical evaluation in the case of multiple measurements and protocols to be defined by the user for output via a printer or PC, individually adjustable for each of the 10 available methods.

  • Accurate measurement of the optical activity of liquids according to ICUMSA, O.I.M.L. And Australian Standard K 157, European and American Pharmacopoeia
  • For the highest demands in quality and research laboratories
  • High precision over the entire measuring range by direct measurement of the angle
  • Short measuring times, no mechanical wear
  • Suitable for microtubes
  • Can be used for various applications with 5 wavelengths; Standard wavelengths are: 365nm, 435nm, 546nm, 579nm and 589nm (others on request) 
  • Automatic temperature correction when using a corresponding tube
  • Calibration-free by spectral lamps 
  • Safe and documented results through fully automatic and digital, objective and fast measurement
  • 10 methods and 10 scales programmable 
  • Continuous measurement without START and RESET
  • Display of optical density 
  • Determination of the optical rotatory dispersion (ORD), measurement at several wavelengths 
  • RS232 to Ethernet adapter for data transfer to the PC 
  • Extensive control software for test routines and configuration of the device 
  • GLP / GMP compliant documentation 
  • 21 CFR part 11 ready, corresponding software “Aquisys 2008” optionally available
  • Robust metal housing 
  • High quality “Made in Germany” as well as advice and service on site possible 
  • Internal measured value memory for 999 values

Body 2

Main specifications

Measurement scales °Optical rotation, °Specific rotation, °Z International Sugar Scale, % Concentration (g/mL, g/100mL, g/L)
7 scales freely
Measuring range ±89° (4 quadrants)
Resolution 0.0001° / 0.01°Z
Precision 0.002° / ± 0.01°Z
Reproducibility 0.001° / ± 0.01°Z
Sensitivity Up to OD 3
Wavelength Up to 5 wavelengths fixed: 365, 435, 546, 578, 589 (others on request)
Response time ≤ 4 sec. over the entire measuring range

Temperature

Temperature measurement If temperature sensor with appropriate tubes is used / T-Cell Peltier tubes
Temperature sample control with appropriate tubes via external water bath / Peltier temperature controlled sample tubes
Temperature control range depends on the water bath / 18 - 40 °C
Temperature resolution depends on the water bath / 0.01 °C
Temperature precision depends on the water bath / ± 0.03 °C
Temperature reproducibility depends on the water bath / 0.05 °C

Additional information

Light source Spectral lamps, calibration free
Display LCD display, monochrom
Operation Alpha numerical keybord, 16 x 16 characters
Operation TOUCH version Alpha numerical keyboard
Interface / Communication 2x RS232, 1x parallel, 1 PS2, USB (adapter)

Conformity

Conformity nternational Pharmacopoeia, OIML, ASTM, ICUMSA, Australian Standard K157